X-ray room requirements in India

What AERB asks for, in plain words


Before a medical X-ray machine can be used on patients in India, the Atomic Energy Regulatory Board (AERB) must license it, and the room must be built to AERB's layout and shielding rules. This guide sums up what AERB's own documents say. Always check the current documents on AERB's website before you build.

Door lining
1.7 mm Pb
Operator barrier
1.5 mm Pb
Aprons
At least 0.25 mm Pb
Licence
Through AERB e-LORA

The licence, step by step

From AERB's e-LORA guidelines for diagnostic radiology (14 July 2016). The portal itself carries the current version.

  1. Register your institute on e-LORA and declare your staff. Every unit needs a qualified operator (an X-ray technologist or radiographer, or a medical practitioner). CT and interventional radiology also need a Radiological Safety Officer (RSO); for other equipment the registrant may designate himself as RSO if no approved RSO is available.
  2. Prepare the room layout in AERB's format (see below), signed by you and authenticated by the supplier.
  3. Apply for procurement permission for new equipment through e-LORA, and buy type-approved equipment from an authorised supplier.
  4. Installation report: after supply, your supplier submits the installation report, including quality-assurance tests and a radiation survey, on behalf of your institute.
  5. Apply for the Licence for Operation once AERB accepts the installation report, within 6 months. After 6 months a fresh QA report is needed.

AERB charges no licence or processing fee. Sources: e-LORA guidelines, sections 2, 7, 8 and 16; AERB/RF-MED/SC-3 (Rev. 2).

Room layout

From AERB's shielding guidelines and safety code.

  • Size: "an appropriate area to facilitate easy movement of staff and proper patient positioning". No fixed number.
  • Door: preferably one entrance door, with a hydraulic closer so it stays shut during exposures, and overlaps at the joints.
  • Window: if there is one, it should be more than 2 m above the outside floor, or shielded up to 2 m.
  • Chest stand: on the wall opposite the entrance door and the control console.
  • Control console: as far as possible from the X-ray tube, with the operator fully behind a protective barrier with a lead-glass window.
  • Mobile X-ray used as a fixed unit must meet all the rules for a fixed installation.
  • Warning sign: a permanent radiation warning symbol and instructions for pregnant women on the entrance door.
1.5 mfrom table centre2.0 mfrom table centre
Brick23 cm20 cm
Concrete15 cm12 cm
Steel2.3 cm2.0 cm
Lead0.17 cm0.15 cm
Wall shielding for radiography and fluoroscopy rooms, from AERB's shielding guidelines. Shielding must reach at least 2 m above the outside floor; 6 to 8 inch concrete is adequate for floor and ceiling. Whether the shielding is enough is finally judged by a radiation survey at commissioning, carried out by the equipment supplier.

Protective devices

Minimum lead equivalence, from AERB/RF-MED/SC-3 Appendix II.

Minimumlead equivalence
Operator barrier (fixed or mobile)1.5 mm Pb
Lead-glass viewing window1.5 mm Pb
X-ray room doors1.7 mm Pb lining
Couch-hanging flaps0.5 mm Pb
Ceiling-suspended glass0.5 mm Pb
Aprons0.25 mm Pb
Gloves0.25 mm Pb
Thyroid shield0.25 mm Pb
Goggles0.25 mm Pb
Gonad shield0.25 mm Pb
Mammography is lighter: its operator barrier and viewing glass need 0.25 mm Pb. Source: AERB/RF-MED/SC-3 (Rev. 2), Appendix II.

Bharat Rays makes aprons, thyroid guards, gloves and gonad shields in 0.25 and 0.5 mm Pb, and protection barriers, screens and lead curtains. Ask us for the lead equivalence of the pieces you need.

Day to day

What AERB's safety code (section 7.2) asks of the operator.

  • Work from behind a protective barrier: a wall, a control room or a mobile protective barrier.
  • Wear a protective apron when using mobile and portable X-ray equipment, and keep at least 2 m from it during exposure.
  • Keep the X-ray room door closed during exposure, and don't let patients, relatives or staff crowd the room.
  • Don't routinely hold patients without aprons; give escorts aprons when they must hold a child or an infirm patient.
  • Wear personnel monitoring devices as your RSO and AERB direct.

This guide summarises AERB's documents for convenience and is not legal advice. AERB's current documents and the e-LORA portal are the authority.

Questions buyers ask

Short, plain answers. Anything else: call 1800 309 9729, toll free.

Is there a minimum size for an X-ray room in India?

AERB's safety code does not set a fixed floor area. It says the room "shall have an appropriate area to facilitate easy movement of staff and proper patient positioning", and the shielding guideline asks you to measure every wall's distance from the centre of the couch, because the shielding needed depends on that distance. See AERB/RF-MED/SC-3, clause 5.4.1.1.

How thick should the walls of an X-ray room be?

For radiography and fluoroscopy, AERB's shielding guideline gives: at 1.5 m from the centre of the patient table, 23 cm brick, 15 cm concrete or 0.17 cm (1.7 mm) lead; at 2.0 m, 20 cm brick, 12 cm concrete or 0.15 cm lead. The shielding must reach at least 2 m above the outside floor, and a 6 to 8 inch concrete floor and ceiling is adequate. See AERB shielding guidelines.

What lead lining does an X-ray room door need?

All doors to an X-ray room need a lead (equivalent) lining of 1.7 mm, according to AERB's safety code (Appendix II.1.5). AERB's shielding guideline also asks for a door closer so the door stays shut during exposures, and overlaps at the joints.

Do I need an AERB licence for an X-ray machine?

Yes. AERB's e-LORA guidelines say it is mandatory for all users of medical diagnostic X-ray equipment to obtain a Licence for Operation, through the e-LORA portal. AERB states that it charges no licence or processing fee.