300 mA
Column and table
- Format
- Column, table and console
- Rating
- 300 mA, 125 kVp, 24 kW
- Tube stand
- Fixed and ceiling-free
Bharat Rays
X-ray machines · line frequency
Column · table · console
The larger line-frequency sets, for an X-ray department: a tube on a vertical column over a radiography table, worked from a separate console. The 300 mA is rated 300 mA at 125 kVp, 24 kW, with a rotating-anode tube, and runs on 440 V.


Copied from the printed Bharat Rays 300 mA sheet.
| 300 mALine frequency | |
|---|---|
| Radiography rating | 300 mA, 125 kVp |
| Output | 24 kW |
| Rectification | Full-wave rectified |
| Timer | 0.06 to 5 s |
| Display | Digital display |
| X-ray tube | Rotating anode |
| Power supply | 440 V |
| Protection | Electronic overload protection |
| Tube stand | Fixed and ceiling-free |
Shown: the 300 mA, from a site photograph.

Column and table

300 mA control
Short, plain answers. Anything else: call 1800 309 9729, toll free.
A 440 V supply (the 300 mA sheet specifies 440 V), space for the column, table and console, and a room built to AERB's layout and shielding guidelines, with a protective barrier for the operator and a lead-lined door. See radiation protection and ask us to check your room plan.
A line-frequency generator works its high-voltage transformer at the mains frequency (50 Hz in India) and rectifies the output, so the tube voltage rises and falls with each mains cycle. A high-frequency generator first converts the mains into a much higher frequency, which gives a nearly steady tube voltage from a smaller transformer and more consistent exposures. In the Bharat Rays range the 10 mA portable and the 100, 300 and 500 mA sets are line frequency; the 2 mA handheld, the 5 mA portable and the 5 to 40 kW machines are high frequency.
Yes. The Atomic Energy Regulatory Board (AERB) requires every user of medical diagnostic X-ray equipment to hold a licence for operation, applied for through its e-LORA portal. The equipment must be type-approved and bought from an authorised supplier, and the room must follow AERB's layout and shielding guidelines. See AERB's e-LORA guidelines for diagnostic radiology, and ask us about the model you are buying.